High Level Anesthesia System Cwm-301d

Product Details
Customization: Available
Type: Anaesthesia Machine
Kind: Gynaecology
Manufacturer/Factory & Trading Company

360° Virtual Tour

Diamond Member Since 2011

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Number of Employees
114
Year of Establishment
2001-04-12
  • High Level Anesthesia System Cwm-301d
  • High Level Anesthesia System Cwm-301d
  • High Level Anesthesia System Cwm-301d
  • High Level Anesthesia System Cwm-301d
  • High Level Anesthesia System Cwm-301d
  • High Level Anesthesia System Cwm-301d
Find Similar Products

Basic Info.

Model NO.
CWM-301D
Certification
CE, ISO13485
Group
Adult
LCD Display
With LCD Display
Transport Package
Standard Export Packing
Specification
ISO, CE
Trademark
CW
Origin
China
HS Code
90189070
Production Capacity
5000PCS/Year

Product Description

Physical specifications
Demension:70x85x140cm,weight:65kg,screen:10.1 LCD
display graphics:wave forms P-T,F-T,V-T ,
loops:P-V,V-F,F-P

working specifications
power:AC110V-240V,50HZ/60HZ, battery power:≥180 mins
Calibration:auto
Pipeline supply gas:O2,N2O

ventilator specifications
patient range:adult,peadiatric
ventilation mode:VCV,A/C,SIMV,MANUAL,STAND-BY,PCV

ventilator parameter
working type:electronic control ,pneumatic driven
tidal volume:20-1500ml ,pressure range:5-60 cmH2O
RRR:1-100bpm,, I:E range:4:1-1:8
Ptrigger range:-20-0 cmH2O,,
minute volume range:>18L/min

System 
Hypoxic guard system:N2O cut off valve,
O2 concentration>25%, safety valve:<12.5kPa
O2 flush:25-75L/min
flow meter:4 tubes cascade 0-15L/min O2,0-15L/min N2O,
volume of CO2 absorber:2L , vaporizer: 1/2 piece 

Monitoring:VT,MV,BPM,Paw,LUNG compliance,inspiratory platform, FiO2

PEEP
type:intergrated electronic control ,
range:off/ 3-30cmH2O
  1. Transferring fresh gas (common gas) containing or not containing anesthetic gas to patient;
  2. Absorbing the CO2 that is exhaled by the patient and transferring the generated O2 to the patient;
  3. Providing a passage to control the patient's inhaling and exhaling;
  4. Exhausting the excessive gas and waste gas.
 

4.6.2 Structure

 

Anesthesia ventilation system mainly comprises: Breathing Bellows; inhalation and exhalation flappers; CO2absorbent; "MC-AC" selection valve; "APL" valve(adjustable pressure limiting valve, exhaust valve); excessive gas discharge port etc.
































1 Breathing Bellows           2 "APL" valve(adjustable pressure limit valve)knob
3Inhaling gas connector 4Exhaling gas connector
5Manual control breathing gas storage bag connector 6 CO2absorbent
7Oxygen concentration monitoring port 8Exhalation flapper
9Inhalation flapper     10"MC-AC" selection knob

Fig 5Anesthesia ventilation system

 

Inhalation flapper and exhalation flapper

 

Inhalation flapper (1) and exhalation flapper (2) are one-way valves controlling direction of respiration gas flows.
These flappers consist of: flapper hood (1),(2), flapper disk (3),(4)
With respiratory movement of patient, inhalation flapper disk and exhalation flapper disk open and close alternative, when the patient.

 
  • nhales, inhalation flapper disk opens and exhalation flapper disk closes, so that gas output by breathing circuit system passes breathing pipeline and enters patient respiratory tract;
  • exhales, inhalation flapper disk closes and exhalation flapper disk opens, so that gas exhaled by patient passes breathing pipeline and enters breathing circuit system.
 

4.6.3 CO2absorbent(cycle absorbent assembly)

 
  • Functions
CO2absorbent contains CO2 absorbent. The absorbent can absorb the patient's exhaled gas, and transfer the generated O2 to the patient.
  • Structure
CO2 absorbent is installed at bottom of the anesthesia ventilation system, composed of absorbent housing, screen, and internal sleeve. The capacity of this equipment CO2 absorbent is approx. 2000mL. Capacity of CO2absorbent of this equipment is about 2000mL.
















1.Screen
 2.Absorbent     
3.Internal sleeve
4.Gas flow direction

Fig 6CO2absorber structure


 

"MC-AC" selection knob

 

To select manual ventilation (bag) or mechanical ventilation (ventilator). When
  • "MC-AC" selection knob is turned to "   ", the patient's breathing is controlled manually;
  • "MC-AC" selection knob is turned to "   ", the patient's breathing is controlled by the ventilator.

"APL"valve (adjustable pressure limit valve, exhaust valve)

"APL" valve is used for adjusting the pressure limit value of the anesthesia ventilation system during manually controlled ventilation. When the pressure in the system exceeds the limit value, the pressure is relieved by the excessive gas exhaust port.
  • Turn clockwise of the "APL" valve knob increases the pressure limit value of the respiratory system. When turning clockwise to the end, the valve closes, and the pressure limit value is approx. 7.0kPa then;
  • Turn counterclockwise of the "APL" valve knob decreases the pressure limit value of the respiratory system. When turning counterclockwise to the end, the "APL" valve opens completely, and the pressure limit value is approx. 0.04kPa then, gas in the respiratory system is exhausted directly through excessive gas exhaust port.
 

 

High Level Anesthesia System Cwm-301dHigh Level Anesthesia System Cwm-301dHigh Level Anesthesia System Cwm-301dHigh Level Anesthesia System Cwm-301dHigh Level Anesthesia System Cwm-301d

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now