★Transport ventialtor
*5.77" screen
SAFETY ,AMBULANCE PARTERNER,ACCURATE DELIVERY ,CONVENIENCE AND EASY CONTROL.
*Ventilation Mode: IPPV.SPONT,A/C,SIMV,SIGH,MANUAL
ventilation setting range:
VT:50-1500ml
RR:4-100/min
IE range: 4:1/1:8
FIO2:45-100Vol %
Monitoring interace:
Vti,Mv,breathing rate,airway pressure
alarms:
No Vt,MV too high(low),airway pressure too high(low),AC power off,back-up battery low.
Power:battery,A/C power 220V/50Hz
Car charger:12V
Battery :≥4 hrs
Net weight:main body <4kg
General
CWH-2010 ventilator is a type of portable electrically and pneumatically controlled synchronous respirator with time switch. It is mainly used for the saving of patients with respiratory failure field surgical operation and resuscitation after surgical operation. It can be used in the ambulance transportation and surgical operation outside, respiration medicine and operation rooms of various types hospitals.
Performance
- Electrical and pneumatic control.
- Time switch, Volume control and Pressure limit ventilation mode.
- Used for adult and child.
- LCD displays the parameter, and can be pre-set the parameter and waveform.
- Ventilation modes:
--IPPV
--SIGH
--SIPPV(A-C)
--SIMV
--SPONT
--MANUAL
- Pre-set ventilation parameters
--Tidal volume
--Frequency
--Sigh time
--I/E
--O2 concentration
--Pressure limitation
--Inspiratory time
--Inspiratory trigger pressure
- Alarm:
--No tidal volume
--Ventilation volume too high (low)
--Airway pressure too high (low)
--The voltage of the battery is too low
- Monitoring parameters:
--Tidal volume
--Minute ventilation volume
--Frequency
--Airway pressure
--SPONT inspiratory trigger indicates
--Inspiratory time
--Respiratory state indicates
Main Technical Parameters:
1, Environment condition
-
- Storage and Transportation conditions:
Ambient temperature: -20°C~+50°C
Relative humidity ≤95%
Atmospheric pressure 70~110kPa
-
- Working conditions:
Ambient temperature: -18°C~+50°C
Relative humidity 15%~95%
Atmospheric pressure 70~110kPa
2. Power source:
--AC ~220V±10%, 50Hz±2%. Input power: 26VA.
--Outside connect DC AC/DC convertor: AC 220V/ DC 12V 0.9A
DC carried in the car: 12V.
--Internal back up battery 12V Working time: ≥3 hours
3, Gas source:
-- gas source: Compressed air (refined) or Oxygen
The given pressure is 0.4Mpa (range.o.25~0.6Mpa)
-- Oxygen cylinder 3L( MAX pressure 15Mpa)
Gas source continous working time ≥40 minutes
4, Basic ventilation parameters:
--Tidal volume(VT) 0.08~1.5 L
--Minute volume 2L~18L
--Frequency (Freq) 4~80 /min (SIMV 1~12 /min)
--I/E 2:1~1:5(SIMV is excepted)
--Inspiratory time(TI) 0.2~6s(when in SIMV)
--Inspiratory trigger pressure (Ptr) -2.0~0 kPa
--Sighs (SIGH) 1~5 times (every 100 times)
--Inspiratory oxygen concentration (FiO2) 50~100%
--Airway spill safety valve £6 kPa
5, Ventilation parameter monitoring range:
--Tidal volume(VT) 0.08~1.5L Tolerance±20 %;
--Minute ventilation volume (Mv) 0~40 L/min Tolerance±20 %;
--Frequency(Freq) 1~80 /min Tolerance±15 %;
--Inhaling time(TI) 0~9.9 s Tolerance±15 %;
--Airway pressure(Ppeak) 0~6 kPa Tolerance±15 %.
6, Alarm and alarm parameters setting:
-- Minute ventilation volume too high (low) (MV High)(MV Low)
When the ventilation volume value is higher (lower) than the set "Ventilation volume too high (low)" value, there is sound alarm and the light illuminates.
Range of "Ventilation volume too high" 5~20 L/min
Range of "Ventilation volume too low" 0~10 L/min
--Airway PIP too high (low) (Ppeak High)(Ppeak Low)
When the airway PIP is higher than the set "Pressure too high" value, there is sound alarm and the light illuminates.
When the airway PIP is lower than the set "Pressure too low" value for 5~20 s, there is sound alarm and the light illuminates.
Range of "Pressure too high" 2~6 kPa
Range of "Pressure too low" 0.4~2 kPa
--No tidal volume(VT No)
When there is no tidal volume output for 6 s, there is sound alarm and the light illuminates
--The pressure of the internal power source (backup battery) is too low,
When the pressure of the internal power source (backup battery) is too low, "The battery pressure is too low" ( ) there is light illuminates alarm
--Max safe pressure:
When the airway pressure is beyond the value of the airway spill safe valve, the valve will expel the gas, the pressure of the Airway spill safe valve: £6Kpa.
--Pressure limit:
When the airway pressure is beyond the given "Pressure is too high", it will release the pressure.
--Alarm persistent time: more than 120s.
--Alarm silence:
If you press the "alarm silence" key, alarm will be paused. But the indication is still on the screen and the light is also twinkling.
Alarm silence time: less than 120s.
7, Default ventilation parameters:
IPPV
--Frequency (Freq) 10/min
--I/E: 1:2
SIPPV
--Frequency (Freq) 10/min
--I/E: 1:2
--Inspiratory trigger pressure (Ptr) -0.1kpa
SIMV
--Frequency (Freq) 10/min
--Inspiratory time(TI) 1s
--Inspiratory trigger pressure (Ptr) -0.1kpa
SIGH
--Frequency (Freq) 10/min
--I/E: 1:2
--time 1/100
8, Default alarm parameter:
--Ventilation volume is too high 10L/min
--Ventilation volume is too low 2.0L/min
--The pressure is too high 4.0kpa
--The pressure is too low 0.5kpa