1.1 Patient range: Support adult and pediatric; 1.2 Display: ≥12 inch high definition TFT colour screen (touchable), can display 3 types of waveform and loops in same screen; 1.3 Power: AC 100-242V, 50/60Hz, with back-up battery, can be used for more than 120 mins after fully charge; 1.4 Working type: Pneumatic driven,electronically controlled; 1.5 Pipeline input range: 0.28~0.6MPa; 1.6 Apply with Day/Night mode,Standby; 1.7 One-key lock for screen, avoid the mis-operation;
2. Ventilation mode and typical function 2.1 Standard ventilation mode: VCV, PCV, PSV, SIMV, APNEA (A/C), CPAP, SIMV/PSV, Bip, APRV,Manual mode etc; 2.2 Apply with Invasive and Non-invasive function; 2.3 Apply with pressure and flow trigger function; 2.4 Apply with Synchronous nebulization and humidify function. High flow oxygen remedy function; 3. Ventilator parameter 3.1 Tidal volume range: 20-2000ml 3.2 Breathing rate: 1-100bpm (1-40 bpm under SIMV); 3.3 Inspiration time: 0.2-12s 3.4 Pressure control/Pressure support:0~80cmH2O 3.5 Inspiration plateau: OFF, 1~50% 3.6 SIGH: OFF, 1~5/100 3.7 Electronic PEEP: OFF,3~40 cmH2O 3.8 CPAP:4~20cmH2O 3.9 O2 concentration: 21%~100%, continuously adjustable 3.10 Inspiration trigger sensitivity: Pressure: OFF, -20~0cmH2O (based on PEEP) Flow:0.5~30L/min
4. Monitoring parameter 4.1 Self-test when start-up the machine: include pressure, flow, O2 sensor etc 4.2 Parameter monitoring: -Tidal volume monitoring: VT for inhalation and exhalation; -Pressure monitoring: Ppeak, Pmean, pressure control level,pressure support level ,PEEP, Pplateau, Peak Inspiratory Negative pressure ; -Breathing rate monitoring: Total breathing rate, breathing rate for spontaneous breath, breathing rate for mechanical ventilation; -Volume monitoring: minute volume, minute volume for spontaneous breath, minute leakage volume; -Flow monitoring: maximum inspiration flow, maximum expiration flow; -Resistance monitoring: inspiration resistance, expiration resistance; -Gas monitoring: O2 concentration; -Special monitoring: Flow trigger depth, Pressure trigger depth, Dynamic Compliance, Static compliance, RVR(RSBI);
4.4 Alarm: VT, MV , airway pressure, continuous pressure, breathing frequency, no connection of tubing, Apnea, O2 concentration,PEEP, O2 or Air failure, AC power failure, back-up battery low, machine fault, etc. Comprehensive 3 level alarm function include audible (adjustable), visual and flicking alarm.
5. Air compressor