Anesthesia ventilator connection
1.Fan2 .Power switch3.CO2 interface 4. Anesthesia Gas interface 5.Power input
6.Equipotential terminals 7.RJ45 interface8.USB interface 9.COM1 interface 10.VGA interface
11.Speaker
Fig 2 Anesthesia ventilator connection port
4.2.2Gas monitor port
Gas monitor port ,see below fig.:
3. CO2 interface: Monitor the CO2 concentration of patient
4. Anesthesia Gas interface: Monitor the anesthesia gas concentrationandtypesof patient
Gas source connection port
Gas source connection port is in the rear of the machine, see Fig 3
1 O2 input port 2 Alternate O2 input 3 N2O input 4 AIR input
Fig 3Gas source connection port
4.2.3 Signal connect port
Signal connect port, see below fig :
1 Oxygen concentration sampling input port (Optional)
2 Tidal volume (pressure/flow) sampling input port
4.3 Anesthesia ventilation
4.3.1Function
It is used for inhalation anesthesia, breathing control or assistant breathing, monitoring and displaying the ventilation parameters of the patient.
4.3.2 Structure
Anesthesia ventilator consists of: panel operation keys; sampling sensor; respiratory monitoring devices; computer control system; regulation and control mechanism; drive mechanism; display; fault alarm and safety protection devices.
The anesthesia ventilator panel see fig 4
1 Display 2 Operation keys 3 "Shuttle adjustment / confirmation" knob
Fig 4 Anesthesia Ventilator panel
4.4 Flow meter
4.4.1Function
Flowmeter with flow control valve which can control and indication of the supply of the patient's common gas (fresh gas) flow.O2,N2O Proportional control unit can guarantee concentration of O2output≥25 %.
4.4.2 Construe
Model |
Construe |
LX-15 |
4 tubes rotameter. 2 tubes of Oxygen, 2 tubes of N2O, consist of thick and thin tubes. Thin tube for small scale; Thick tube for large scale. Total scale is 15L |
LX-15V |
5 tubes rotameter. 2 tubes of Oxygen, 2 tubes of N2O, 1 for air. Consists of thick and thin tubes. Thin tube for small scale; Thick tube for large scale. Air tube is thick only. Total scale is 15L |
LX-15 |
6 tubes rotameter. 2 tubes of Oxygen, 2 tubes of N2O, 2 tubes for Air. Consists of thick and thin tubes. Thin tube for small scale; Thick tube for large scale. Total scale is 15L |
- Flowmeter is a rota type with O2,N2O,Air. There are thick and thin tubes over 2 tubes flowmeter.
- Flowmeter has a flow control valve, controlling the total portion of the mixed gas of O2,N2O,Air. Flow control valve can be adjusted with anti-clockwise to increase the flow rate, alternatively clockwise to reduce.
- There is proportional control device between O2 and N2O to ensure the O2 concentration in the mixed gas is no less than 25%. N2O will be shut down once O2 is closed.
- The structure of flowmeter as below:(for example LX-15V)
4.4.3 Accuracy
The accuracy of the full scale of 10% or 300 mL/min (whichever is larger) to full scale is between and 10% of the indicated value.
Below the full scale of 10% or 300 mL/min (whichever is the large value) of the accuracy of 4
- The flow meter is equipped with flow control valve, which controls and indicates the flow of common gas (fresh gas) supplied to patient.
- Flow control valve counter-clockwise to increase the flow, clockwise to reduce flow.
- O2, N2O proportion control device can ensure concentration of the output O2 is≥25 %.When O2 closed, N2O automatically shut down.
- O2 flush flow 25 L/min~ 75 L/min
4.5O2 flush button
O2 flush button provide high-flow oxygen to the anesthesia system by the ACGO gas inlet:
- Press the button to input high-flow oxygen;
- Release the button to stop the gas flow.
Physical specifications
Demension:70x85x140cm,weight:65kg,screen:12.1 LCD
display graphics:wave forms P-T,F-T,V-T ,
loops:P-V,V-F,F-P(zoom in/out ,freeze,comparison)
working specifications
power:AC110V-240V,50HZ/60HZ, battery power:≥180 mins
Calibration:auto
Pipeline supply gas:O2,N2O,air
ventilator specifications
patient range:adult,peadiatric
ventilation mode:VCV,A/C,SIMV-V,AIMV-P,MANUAL,STAND-BY,PCV,PSV
ventilator parameter
working type:electronic control ,pneumatic driven
tidal volume:20-1500ml ,pressure range:5-60 cmH2O
RRR:1-100bpm,, I:E range:4:1-1:8
Ptrigger range:-20-20 cmH2O,,
minute volume range:>18L/min
System
Hypoxic guard system:N2O cut off valve,
O2 concentration>25%, safety valve:<12.5kPa
O2 flush:25-75L/min
flow meter:5 tubes cascade 0-15L/min O2,0-15L/min N2O,
0-15L/min AIR
volume of CO2 absorber:2L , vaporizer: 2/2 piece
Monitoring:VT,MV,BPM,Paw,LUNG compliance,inspiratory platform, FiO2
PEEP
type:intergrated electronic control ,
range:off/ 3-30cmH2O