Main features:
10.1" color active matrix TFT display, 1024 x 600 pixels, adjustable brightness
Displays all setting and alarm parameters; waveforms (P-T, F-T, V-T) and loops (P-V, F-V, F-P)
Ventilation modes: VCV, SIMV, A/C, MANUAL, STAND-BY
Applicable for adult and pediatric patients
Integrated Adjustable Pressure Limiting (APL) valve, range 1 - 75 cmH2O
Low-compliance breathing circuit (<= 4 mL/100 Pa) with low inspiratory/expiratory resistance (< 6.0 cmH2O @60 L/min)
Supports up to 2 vaporizers (Selectatec mounting, interlocking function); agents: Halothane, Enflurane, Isoflurane, Sevoflurane
Mechanical O2-N2O link system ensures O2 concentration not lower than 25%
Built-in lead-acid battery (12 Vdc, 4 Ah) provides 180 min backup (new, fully charged @25 C)
Safety design: manual ventilation, gas and agent delivery remain possible during power or battery failure
Main specifications:
Power: 220 - 240 V AC, 50/60 Hz; built-in battery backup 180 min
Gas supply: O2, N2O, Air (optional); pipeline input 0.28 - 0.6 MPa (NIST/DISS); cylinder input optional (PISS, max. 2 cylinders)
Ventilation modes: VCV, A/C, SIMV, MANUAL, STAND-BY
Setting parameters: Tidal volume 20 - 1500 mL, Breath rate 1 - 100 bpm, I/E ratio 4:1 - 1:8, Trigger pressure -20 - 0 cmH2O, Pressure limit 10 - 100 cmH2O
Monitoring parameters: Tidal volume (0 - 2500 mL), Minute volume (0 - 100 L/min), FiO2 (15% - 100%), Peak/Mean airway pressure (-20 - 100 cmH2O), Breath rate (0 - 100 bpm), I/E ratio
Waveform display: Pressure-time, Flow-time, Volume-time
Loop display: Pressure-Volume, Flow-Volume, Flow-Pressure
Alarm: Minute volume high/low, Peak airway pressure high/low, FiO2 high/low, CO2 high/low (optional), Supply failure, AC power failure, low voltage; alarm silence countdown 120 - 0 s
Vaporizer: VPI type, max. 2 units, Selectatec mount with interlock
Breathing system: Compliance <= 4 mL/100 Pa; Inspiration/Expiration resistance < 6.0 cmH2O @60 L/min
Physical specifications: Height 1400 mm, Width 850 mm (incl. breathing system), Depth 700 mm, Weight < 65 kg (without vaporizers & cylinders)
Environmental: Operating temperature 5 - 40 C, humidity <= 80%; Storage temperature -10 - 40 C (main unit), humidity <= 90%